TD-150-BLE: Difference between revisions
→Installation of the configurator and connection to the sensor
| Line 417: | Line 417: | ||
== '''<big>Installation of the configurator and connection to the sensor</big>''' == | == '''<big>Installation of the configurator and connection to the sensor</big>''' == | ||
The sensor can be configured using [https://www.fmeter.ru/download/_ftp/all/dut/FLS_Configurator.zip?v=210521143455 '''<big>конфигуратора на ПК</big>'''] (From here onwards- "'''configurator'''"). | |||
''' | '''Connect the sensor to the USB-RS-485 converter using a 6-pin MOLEX connector''' or using cable clamps if a cable route is connected to the sensor. Orange wire is line A of the RS-485 sensor interface, white wire is line B of the RS-485 sensor interface, black wire is GND, red is PWR. | ||
[[Файл:Датчик подключенный при помощи MOLEX.png|без|мини|800x800пкс|'''<big>Датчик подключенный при помощи MOLEX</big>''']] | [[Файл:Датчик подключенный при помощи MOLEX.png|без|мини|800x800пкс|'''<big>Датчик подключенный при помощи MOLEX</big>''']] | ||
[[Файл:Датчик подключенный при помощи кабель-трассы (укороченной) и кабельных зажимов.png|без|мини|800x800пкс|'''<big>Датчик подключенный при помощи кабель-трассы (укороченной) и кабельных зажимов</big>''']] | [[Файл:Датчик подключенный при помощи кабель-трассы (укороченной) и кабельных зажимов.png|без|мини|800x800пкс|'''<big>Датчик подключенный при помощи кабель-трассы (укороченной) и кабельных зажимов</big>''']] | ||
We recommend using our Escort C200M/C200M2 USB-RS-485 converter, since we cannot guarantee 100% compatibility of our devices with converters from other brands. | |||
When working with a laptop, we recommend connecting it to the power supply and/or connecting an additional USB cable to the ADD connector of the PWR C200M. Otherwise, there may not be enough power to operate the sensor and transmitter. | |||
Along with installing the configurator 1.0.2.38, the drivers for the C200M will be installed automatically. | |||
If you use C200M2 on Windows 10 and 11 operating systems, drivers should be installed automatically from Windows Update, on the Windows 7 operating system and below, you may need to[https://remontka.pro/disable-drivers-signature-check-windows-10/ отключение электронной подписи драйверов] and manually install [https://www.fmeter.ru/download/_ftp/escort_c-200m/%D0%94%D1%80%D0%B0%D0%B9%D0%B2%D0%B5%D1%80%20%D0%B4%D0%BB%D1%8F%20C-200M2.zip?v=150323104902 драйверов на С200М2]. | |||
If the driver was installed correctly, then after connecting the converter to your PC/laptop, you will see the STMicroelectronics Virtual COM Port ('''1''')(C200M) or USB-SERIAL CH341A ('''2''')(C200M2) device in the COM and LPT ports section of the Windows device manager " to enter this menu, press '''win+r''' and enter '''devmgmt.msc''' and press '''OK''' ('''3''') and then expand the com ports submenu ('''4''')" | |||
The com port number displayed in this menu is also needed to connect the sensor. | |||
[[Файл:Ком порты С200М и С200М2.png|900x900пкс]] | [[Файл:Ком порты С200М и С200М2.png|900x900пкс]] | ||
After connecting the converter, sensor to it and checking the installation of drivers by checking the com port number of the converter, you need to open the configurator, select the desired com port which we could find in the device manager ('''1''') and press the '''FLS''' button ('''2'''). | |||
[[Файл:Выбор ком порта и ДУТ.png|750x750пкс]]<blockquote> | [[Файл:Выбор ком порта и ДУТ.png|750x750пкс]]<blockquote>The connection to the FLS should be made '''within 15 seconds after the sensor has been connected''' to power, if the sensor operating mode has been changed from RS-485 to any other. </blockquote>After connecting the sensor you should see this menu: | ||
[[Файл:Описание основного меню конфигуратора на ПК, проводного дут .png]] | [[Файл:Описание основного меню конфигуратора на ПК, проводного дут .png]] | ||
# | # Sensor serial number | ||
# | # Sensor firmware version (FW) | ||
# | # Sensor temperature | ||
# | # Current sensor level | ||
# | # Current CNT level (raw level value) of the sensor | ||
# | # LLS network addresses connected to this converter (if there is more than one address in this list, it means either more than one sensor is connected to the line at the same time or there is interference on the line, in this case it is necessary to check the connection to the FLS for the presence of other conflicting devices and it is necessary to close other programs using the com ports for example tracker configurator) | ||
# | # Network address of the polled sensor (This address is used when connecting in RS485 mode) | ||
# | # Current sensor operating mode | ||
# | # Current range of output values (1-1023 or 1-4095) | ||
# | # Current filtration type and degree | ||
# | # Connected sensor model | ||
== '''<big>Sensor calibration</big>''' == | == '''<big>Sensor calibration</big>''' == | ||