Find the '''required sensor (1)''' by typing the '''last 6 digits of its serial number'''. You can find the serial number on the sensor head.
Find the '''required sensor (1)''' by typing the '''last 6 digits of its serial number'''. You can find the serial number on the sensor head.


Или, просто, выберите требуемый датчик из списка и кликни кнопку '''Подключиться(2)''' . В устройстве '''Android''' вы можете кликнуть на датчик, при этом отобразится пакет данных полученных в режиме '''advertising(3).'''
You can also simply select the required sensor from the list and click the '''Connect (2) button'''. On an '''Android device''', you can click on the sensors name, and a package of data received in '''advertising mode (3)''' will be displayed.


[[File:Connection and advertising TD-150 BLE.png|frameless]]
[[File:Connection and advertising TD-150 BLE.png|frameless]]


После успешного подключения вы увидите главный экран датчика.
Once connected successfully, you will see the main screen of the sensor.


[[Файл:Основные параметры TD-150-BLE.png|1059x1059пкс]]
[[Файл:Основные параметры TD-150-BLE.png|1059x1059пкс]]
Line 155: Line 155:
[[File:TD-150 BLE additional features.png|frameless|711x711px]]
[[File:TD-150 BLE additional features.png|frameless|711x711px]]


Затем по появившейся вкладке в графе по работе с паролем '''введите пароль(1)''', который будет использоваться впоследствии, и нажмите на кнопку '''Установить(2)'''.
Then on the tab that appears, in the password operation box, enter the '''password(1)''' to be used later and press '''Install(2)'''.


[[File:TD-150 BLE setting password .png|frameless]]
[[File:TD-150 BLE setting password .png|frameless]]
Line 168: Line 168:
[[Файл:Калибровка TD-150-BLE.gif]]
[[Файл:Калибровка TD-150-BLE.gif]]


После того, как вы удлинили или укоротили трубки датчика - необходимо провести процедуру калибровки датчика.
After you have lengthened or shortened the sensor tubes, you need to carry out the sensor calibration procedure.


Для этого необходимо:
To do this you need:


* Перейти в меню "'''Настройки'''"
* Go to the '''"Settings"''' menu


[[File:TD-150 BLE settings page.png|frameless]]
[[File:TD-150 BLE settings page.png|frameless]]
Line 179: Line 179:
* Fill the tubes with fuel (by sealing the drain holes with duct tape and filling the tubes, turning the sensor upside down, or submerging the sensor tubes completely in fuel)
* Fill the tubes with fuel (by sealing the drain holes with duct tape and filling the tubes, turning the sensor upside down, or submerging the sensor tubes completely in fuel)
* Wait for the CNT level to stabilize ('''2''')
* Wait for the CNT level to stabilize ('''2''')
* Уберите выделение с ползунка "'''Калибровка без топлива'''" ('''1''')
* Deselect the '''"Calibration without fuel"''' slider ('''1''')
* Нажмите "'''Полный'''" ('''3''')
* Press '''"Full"''' ('''3''')
* Значение "'''Полный'''"('''4''') должно измениться на значение близкое к значению текущего CNT ('''2'''), но не равно ему, так как данное значение задается согласно термокомпенсации датчика
* The value '''"Full"''' ('''4''') should change to a value close to the value of the current CNT ('''2'''), but not equal to it, since this value is set according to the temperature compensation of the sensor


[[File:TD-150 BLE Full calibration.png|frameless]]
[[File:TD-150 BLE Full calibration.png|frameless]]