TD-BLE: Difference between revisions
→Calibration of a tank whose height varies along its length
| Line 384: | Line 384: | ||
# However, before continuing calibration of the second FLS, it is necessary to return the nominal portion volume to the original (i.e. double it). IMPORTANT: The actual portion size still remains unchanged until the tank is finally filled and the calibration process is completed. | # However, before continuing calibration of the second FLS, it is necessary to return the nominal portion volume to the original (i.e. double it). IMPORTANT: The actual portion size still remains unchanged until the tank is finally filled and the calibration process is completed. | ||
Thus, the resulting calculation tables (calibration tables) will be adequately accepted by the monitoring platform if a third FLS (virtual) is created in it, which is the sum of two real FLS.[[File:Tilted FLS 1.png|none|thumb|'''<big>Tank calibration example FLS 1</big>''']] | Thus, the resulting calculation tables (calibration tables) will be adequately accepted by the monitoring platform if a third FLS (virtual) is created in it, which is the sum of two real FLS. | ||
[[File:Tilted FLS 2.png|none|thumb|'''<big>Tank calibration example FLS 2</big>''']] | [[File:Tilted FLS 1.0.png|none|thumb|512x512px|'''<big>Tank calibration example FLS 1</big>''']] | ||
[[File:Tilted FLS 2.png|none|thumb|'''<big>Tank calibration example FLS 2</big>'''|458x458px]] | |||
== '''<big>Calibration of a tank whose height varies along its length</big>''' == | == '''<big>Calibration of a tank whose height varies along its length</big>''' == | ||