# However, before continuing calibration of the second FLS, it is necessary to return the nominal portion volume to the original (i.e. double it). IMPORTANT: The actual portion size still remains unchanged until the tank is finally filled and the calibration process is completed.
# However, before continuing calibration of the second FLS, it is necessary to return the nominal portion volume to the original (i.e. double it). IMPORTANT: The actual portion size still remains unchanged until the tank is finally filled and the calibration process is completed.


Thus, the resulting calculation tables (calibration tables) will be adequately accepted by the monitoring platform if a third FLS (virtual) is created in it, which is the sum of two real FLS.[[File:Tilted FLS 1.png|none|thumb|'''<big>Tank calibration example FLS 1</big>''']]
Thus, the resulting calculation tables (calibration tables) will be adequately accepted by the monitoring platform if a third FLS (virtual) is created in it, which is the sum of two real FLS.
[[File:Tilted FLS 2.png|none|thumb|'''<big>Tank calibration example FLS 2</big>''']]
[[File:Tilted FLS 1.0.png|none|thumb|512x512px|'''<big>Tank calibration example FLS 1</big>''']]
[[File:Tilted FLS 2.png|none|thumb|'''<big>Tank calibration example FLS 2</big>'''|458x458px]]


== '''<big>Calibration of a tank whose height varies along its length</big>''' ==
== '''<big>Calibration of a tank whose height varies along its length</big>''' ==