You should clean the sensor tubes—preferably by blowing them out through the drainage holes using compressed air.
You should clean the sensor tubes—preferably by blowing them out through the drainage holes using compressed air.


If this error occurred after the sensor was put into operation, it is highly likely that contaminants entered the tubes from the reservoir. In this case, it is also necessary to clean the reservoir thoroughly.
If this error occurred after the start of operation of the sensor, it means that most likely these impurities got into the sensor tubes from the container and in this case it is also needed to ensure that the tank itself doesn't contain any contaminations. Clean the container, if necessary. Note that an ALS installed outside a contaminated container may function correctly, but the same sensor installed in such container may generate this error code.