TD-500: Difference between revisions
→Sensor calibration
| Line 152: | Line 152: | ||
* Go to menu '''"Sensor calibration"''' | * Go to menu '''"Sensor calibration"''' | ||
[[ | [[File:TD 150 calibration.png|frameless|634x634px]] | ||
* Insert the centralizer into the tubes | * Insert the centralizer into the tubes | ||
| Line 161: | Line 161: | ||
* The value '''"Full"''' ('''4''') should change to a value close to the value of the current CNT ('''1'''), but not equal to it, since this value is set according to the temperature compensation of the sensor | * The value '''"Full"''' ('''4''') should change to a value close to the value of the current CNT ('''1'''), but not equal to it, since this value is set according to the temperature compensation of the sensor | ||
[[ | [[File:Wired calibration.png|frameless|344x344px]] | ||
[[ | [[File:Centralizer-Centrator.png|none|thumb|509x509px|'''<big>Centralizer</big>''']] | ||
[[ | [[File:Centralizer.png|none|thumb|512x512px|'''<big>Centralizer at the end of the tubes</big>''']] | ||
<gallery widths="350" heights="300"> | <gallery widths="350" heights="300"> | ||
Файл:Закрытие дренажных отверстий, переворачивание датчика и заполнение трубок топливом.png|'''<big>Закрытие дренажных отверстий, переворачивание датчика и заполнение трубок топливом</big>''' | Файл:Закрытие дренажных отверстий, переворачивание датчика и заполнение трубок топливом.png|'''<big>Закрытие дренажных отверстий, переворачивание датчика и заполнение трубок топливом</big>''' | ||