TD-150-BLE: Difference between revisions
→The sensor does not connect or is not recognized in the application
| Line 820: | Line 820: | ||
If the sensor does not connect to the configurator, do the following: | If the sensor does not connect to the configurator, do the following: | ||
* Make sure that the correct COM port number is selected and that drivers and libraries are installed (STMicroelectronics Virtual COM Port (1)(C200M) or USB-SERIAL CH341A (2)(C200M2) in the COM and LPT ports section of the Windows device manager "to enter this menu, press win+r and enter devmgmt.msc and press OK (3) and then expand the com ports submenu (4)" [[ | * Make sure that the correct COM port number is selected and that drivers and libraries are installed (STMicroelectronics Virtual COM Port (1)(C200M) or USB-SERIAL CH341A (2)(C200M2) in the COM and LPT ports section of the Windows device manager "to enter this menu, press win+r and enter devmgmt.msc and press OK (3) and then expand the com ports submenu (4)" | ||
* [[File:Com ports connected.png|frameless|748x748px]] | |||
* [[File:Com ports selection.png|frameless|747x747px]] | |||
* If possible, connect another sensor that is sure to work; if it connects, then there are no problems with the COM port or converter | * If possible, connect another sensor that is sure to work; if it connects, then there are no problems with the COM port or converter | ||