TD-500: Difference between revisions
→Calibration without fuel
| Line 165: | Line 165: | ||
[[File:Centralizer.png|none|thumb|512x512px|'''<big>Centralizer at the end of the tubes</big>''']] | [[File:Centralizer.png|none|thumb|512x512px|'''<big>Centralizer at the end of the tubes</big>''']] | ||
<gallery widths="350" heights="300"> | <gallery widths="350" heights="300"> | ||
File:Closing the drain holes, rotating the sensor, and filling the tubes with fuel.png|'''<big>Closing the drain holes, inverting the sensor and filling the tubes with fuel</big>''' | |||
File:Filling the tubes by immersing the sensor in fuel (drain holes open).png|'''<big>Filling the tubes by immersing the sensor in fuel (drain holes open)</big>''' | |||
</gallery> | </gallery> | ||
| Line 175: | Line 175: | ||
* Click "'''OK"''' | * Click "'''OK"''' | ||
[[ | [[File:PC Empty calibration.png|frameless|366x366px]] | ||
* The sensor level should display as 1, the sensor calibration process is complete. | * The sensor level should display as 1, the sensor calibration process is complete. | ||
| Line 181: | Line 181: | ||
Thus, '''CNT''' should increase as the sensor tubes fill with fuel. It should change from a value close to the '''Empty''' calibration value to the '''Full''' calibration value. | Thus, '''CNT''' should increase as the sensor tubes fill with fuel. It should change from a value close to the '''Empty''' calibration value to the '''Full''' calibration value. | ||
[[ | [[File:TankEmptyFull.png|frameless|512x512px]]<blockquote>'''<big>ATTENTION! UNBLOCK THE DRAINAGE HOLES AFTER CALIBRATION!!!</big>'''</blockquote>[[File:Unblocking drain holes.png|frameless|585x585px]] | ||
== '''<big>Calibration without fuel</big>''' == | == '''<big>Calibration without fuel</big>''' == | ||